We use cookies to make your experience better. To comply with the new e-Privacy directive, we need to ask for your consent to set the cookies.
Minimum Flowrate (ml/min) | US$1,330.90 |
---|---|
Maximum Flowrate (ml/min) | US$6,654.50 |
Brand | SKC |
Available Outside the UK? | No |
Flow Range:
1000 to 5000 ml/min (5 to 500 ml/min requires optional low flow adapter kit)
Compensation Range:
5000 ml/min at 10 inches water back pressure
4000 ml/min at 20 inches water back pressure
2000 ml/min at 50 inches water back pressure
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min) | 1.0 | 1.5 | 2.0 | 2.5 | 3.0 | 3.5 | 4.0 | 5.0 |
Filter/Pore Size (μm) | ||||||||
25-mm MCE/0.8 | 6 | 9 | 12 | 15 | 18 | 21 | 25 | 31 |
25-mm MCE/0.45 | 14 | 22 | 28 | 35 | 40 | 44 | 50 | 63 |
37-mm MCE/0.8 | 2 | 3 | 4 | 5 | 6 | 7 | 9 | 11 |
37-mm PVC/5.0 | 1 | 1 | 2 | 2 | 2.5 | 3 | 3 | 4 |
37-mm, polycarbonate/0.45 | 4 | 6 | 8 | 10 | 12 | 15 | 17 | 21 |
25-mm MCE/0.45 microvacuum | 21 | 31 | 40 | 48 | 59 | 69 | 79 | 100 |
37-mm PTFE/1.0 | 1.5 | 2.5 | 4 | 5.5 | 7 | 8 | 9.5 | 12 |
Flow Compensation System:
Patented* isothermal closed loop flow sensor
Accuracies:
Timing:1 min/mo at 25 C
Flow Rate: ± 5% of set-point after calibration to desired flow
Battery Charge Level Indicator:
Icon displays on LCD at full, mid, low charge, imminent low battery fault, and low battery fault.
Temperature Range:
Operating:32 to 113 F (0 to 45 C)
Charging:32 to 113 F (0 to 45 C)
Storage:-4 to 95 F (-20 to 35 C)
Operating Humidity: 0 to 95% non-condensing
Typical Run Time† :
XR5000 Model | 2 L/min | 5 L/min |
High-power Li-Ion | 40 hrs | 22 hrs |
Standard Li-Ion | 20 hrs | 11 hrs |
† Results of run time tests using 37-mm, 0.8-μm MCE filters with new pumps and batteries. Pump and battery performance may vary.
For extended run times, the pump may be operated while attached to the charger.
Timed Run, Run Delay, and Continuous Run 1 to 9999 minutes (6.8 days). If run time exceeds
Display Range: 6.8 days, timer display rolls over.
Flow Fault:
If the pump is unable to compensate for > 15 seconds due to excessive back pressure, the pump stops, displays flow fault icon, and holds run time display. Auto-restart is attempted every 15 seconds up to 5 times.
Low Battery Fault: 15 seconds to sleep